General
AID Series alarm and display device is remote display and sound and light alarm device by using advanced fieldbus communications technology. AID Series alarm and display device have four models: AID10, AID120, AID150 and AID200. Products are classified according to the installation method and display. The device can be installed on the operating room or information control panel of intensive care unit, the medical staff can easily know the status of medical isolated power systems, and the type of fault when the system fails.
AID150 centralized alarm and display instrument adopts the LCD liquid crystal display and achieves data exchange with AIM-M200 medical intelligent insulation monitoring instrument through CAN communication interface of self-defined protocol, which can real-time monitor multi-channel data of AIM-M200 medical intelligent insulation monitoring instrument.
Function
* Can read and display monitoring data and alarm information of insulation monitor in real-time by bus communication
* Can set the alarm parameters of insulation monitor remotely
* When the system fails, products can generate sound and light alarm, if you press the mute button, the sound can be turned off, but the corresponding fault indicator does not go out until the fault is removed so far.
Specification
Model Parameter |
AID10 |
AID120 |
AID150 |
AID200 |
||
Auxiliary power |
Voltage |
DC24V |
||||
Consumption |
<0.6W |
|||||
Insulation resistance Measurement range |
-- |
0-999KΩ |
-- |
-- |
||
Insulation alarm value range |
-- |
50-999KΩ |
-- |
-- |
||
Transformer load rate display |
-- |
Percentage display |
-- |
-- |
||
Load current alarm value settings |
-- |
14A,18A, 22A, 28A, 35A, 45A |
||||
Temperature alarm value range |
-- |
0-200℃ |
||||
Alarm |
Sound and light alarm |
|||||
Alarm type |
Insulation fault, overload, over-temperature, equipment failure |
|||||
Communication |
RS485 interface, Modbus-RTU protocol |
|||||
Display |
-- |
LED display |
LCD display |
LCD display |
Dimension
Display
Application
Wiring and Installation
Terminal Description
Note: The 24V power supply can be connected by multiple copper wires of 2x1.5mm2, and the RS485 communication terminal can be connected by shielded twisted pair of 2x1.5mm2
Installation
Common Problems
For common problems, the causes can be determined and the faults can be eliminated according to the phenomenon of each instrument and the fault types:
Fault phenomenon |
Possible causes and troubleshooting |
The instrument is not lit. |
The 24V power supply is not connected well. Check the wirings of 24V and G terminals and rewire. |
Communication indicator does not flash |
In the system, the AIM-M monitor communication line is not connected properly. Check the communication line and confirm whether the matching resistance is connected properly. |
Note: If the above faults occur, interrupt the power to troubleshoot, and adjust the wirings until everything is normal
Acrel Medical IT System Insulation Monitoring and Fault Location Products Selection Table V1.0
Note:“□”is optional,“—”is not optional,“■”is selection.
Medical Single Phase Isolation Transformer
AITR-3150
AITR-5000
AITR-6300
AITR-8000
AITR-10000
Capacity
3150VA
5000VA
6300VA
8000VA
10000VA
□
□
□
□
□
Insulation monitor
AIM-M10
AIM-M100
AIM-M200
Power supply
AC220±10%
□
AC85~265V
DC100-300V
□
DC24V
□
Communication
RS485
□
RS485+RS485
□
RS485+CAN
□
Output
DO1(over temperature)
□
DO1(over temperature)
DO2(alarm)
□
DO1(over temperature)
DO2(alarm) DO3(programmable)
□
1 channel DC24V
□
Support fault location
No
□
□
Yes
□
Power supply
ACLP10-24
HDR-60-24
Insulation monitor
AIM-M100
□
AIM-M200
□
Alarm indicator
AID10
AID120
AID150
AID200
Display
LED
□
□
LCD
□
□
Multi-machine display
No
□
□
Yes
□
□
Current transformer
AKH-0.66P26
■
Test signal generator
ASG150
Support fault location
No
—
Yes
■
Fault locator
AIL150-4
AIL150-8
Support fault location
No
—
—
Yes
□
□
Locate number of channels
<=4
□
>4; <8
□
>8 Note 1
□
□
Other
Note: 1. More than 8 channels can be freely combined, and the maximum support 24 locate channels.
Ordering example
Example 1: Isolation transformer 8kVA; AC220V; Insulation monitoring; Operating room LCD display; DO output programmable,According to the selection table, confirm the model is AITR-8000, AIM-M100, ACLP10-24, AKH-0.66P26, AID150.
Example 2: Isolation transformer 10kVA; Insulation monitoring; Fault location; ICU LCD display; DO output; Locate 12 loops;
According to the selection table, confirm the model is AITR-10000, AIM-M200, HDR-60-24, AKH-0.66P26, AID150, ASG150, AIL150-4, AIL150-8