● Insulation monitoring for medical IT systems

General
AIM-M200 medical intelligent insulation monitoring instrument adopts advanced microcontroller technology, which has high integration, compact size, convenient installation and integrates intelligence, digitalization and networking in one. It is ideal selection for insulation monitoring of isolation power system in Class 2 medical locations such as operating room and intensive care unit.
Function
1) Real time monitor and fault alarm of the ground insulation resistance,
2) Transformer load current and transformer winding temperature of the monitored IT system,
3) Be used with insulation fault locator, remotely starting fault-locating and displaying locating results when there are insulation faults,
4) Real-time monitor the line disconnection fault, temperature sensor disconnection fault and the functional grounding line disconnection fault of the monitored system, and give the alarm indication within 2S after the fault occurs,
5) Relay alarm output, LED alarm indication and other faults indication functions,
6) Used for centralized alarm and display instrument, test signal generator, insulation fault locator and upper computer management software communications, and can monitor the operation status of IT system in real time.
7) Events logging function, which can record alarm occurrence time and fault type and is convenient for operation personnel to analyze the operation conditions of system and promptly eliminate the faults.
Specification
Product Selection
Wiring
Display
Dimension
Typical Connection
Installation Size of AKH-0.66P26 is shown in the following figure:
LED Indicator Instruction
Indicator status |
Instruction |
On |
When the device is in normal operation, the indicator flashes, and the flickering frequency is about once a second. |
Comm |
Indicate communication status, when there is data transmission, the indicator flashes. |
Insulation |
When the insulation resistance exceeds the alarm setting value, or when FK/LL is disconnected, the indicator flashes to alarm. |
Overload |
When the system load current exceeds the alarm setting value, the indicator flashes to alarm |
Overtemp |
When the transformer temperature exceeds the alarm setting value or the temperature sensor wiring is disconnected, the indicator flashes to alarm. |
Data Setting
First Menu |
Second Menu |
Range and Description |
Set Comm |
485 Ad |
1-247 |
Baud |
4800,9600,19200 |
|
Can Ad |
1-110 |
|
LOCAT |
YES,NO |
|
Set Alarm |
Res |
50-999 |
Irms |
14,18,22,28,35,45 |
|
Temp |
0-200 |
|
Set Time |
Date
|
Year-month-day |
Time |
Hour-minute-second |
|
Set Others |
Pass
|
0000-9999 |
Contra |
0-63 |
|
BL |
1-99 |
|
ClrSOE |
Clear the fault record |
Common Problem
Make sure the connection is correct before the system is powered on. For common problems, according to the fault phenomena in the table below, targeted investigation can be carried out.
Fault phenomenon |
Possible causes and troubleshooting |
LCD: LL open, the Insulation indicator is lit. |
Terminal 4 and terminal 5 of AIM-M200 are not reliably connected to L1 and L2 of the IT system. Check the connection and make sure it is reliable. |
LCD: FK open, the Insulation indicator is lit. |
Terminal 13 and terminal 14 of AIM-M200 are not reliably connected to LEB. Check the connection and make sure it is reliable. |
LCD: TC open, the Overtemp indicator is lit. |
Terminal 11 and terminal 12 of AIM-M200 are not reliably connected to the two ST terminals of the isolation transformer. Check the connection and make sure it is correct. |
LCD: insulation fault, the Insulation indicator is lit. |
At least one of the L1 and L2 wires in IT system has grounding fault, which can be restored to normal after removal. |
The device can't light up. |
The working power supply of AIM-M200 is not connected properly. Check the connection of terminals 1 and 2 and make sure it is correct |
Note: All the above faults should be checked by power off, and the wiring should be adjusted until everything is normal.
AUX Power |
Voltage |
DC 18-36V |
Temperature monitoring |
Thermal resistor |
2 Pt100 |
Power consumption |
≤3VA |
Measuring range |
-50~+200℃ |
||
Insulation monitoring |
Resistance measuring range |
10-999kΩ |
Alarm value range |
0~+200℃ |
|
Response value |
50-999kΩ |
Alarm output |
Output mode |
2 Relays |
|
Relative uncertainty |
±10% |
Contact rating |
AC 250V/3A DC 30V/3A |
||
Response time |
≤3s |
Environment |
Operating temperature |
-10~+55℃ |
|
Permissible system leakage capacitance Ce |
≤5uF |
Transport temperature |
-25~+70℃ |
||
Measuring voltage Um |
≤12V |
Storage temperature |
-25~+70℃ |
||
Measuring current Im |
≤50uA |
Relative humidity |
5%-95%,No condensation |
||
Impedance Zi |
≥200kΩ |
Altitude |
≤2500m |
||
Internal DC resistance Ri |
≥240kΩ |
IP degree |
IP30 |
||
Permissible extraneous DC voltage Ufg |
≤DC280V |
Rated impulse voltage / pollution degree |
4KV/Ⅲ |
||
Load current monitoring |
Measuring Value |
2.1-50A |
EMC |
IEC 61326-2-4 |
|
Alarm Value |
5-50A |
Communication |
CAN |
||
Measuring accuracy |
±5% |
RS485(Modbus-RTU) |